Comparison of CAD/CAM manufactured implant-supported crowns with different analyses
| dc.authorid | 0000-0003-1194-0703 | |
| dc.authorid | 0000-0002-2865-372X | |
| dc.contributor.author | Yegin, Elif | |
| dc.contributor.author | Atala, Mustafa Hayati | |
| dc.date.accessioned | 2025-05-10T19:34:27Z | |
| dc.date.issued | 2020 | |
| dc.department | İstanbul Medeniyet Üniversitesi | |
| dc.description.abstract | Background Present study compared the failure load of CAD/CAM-manufactured implant-supported crowns and the stress distribution on the prosthesis-implant-bone complex with different restoration techniques. Methods The materials were divided into four groups: group L-M: lithium disilicate ceramic (LDS, monolithic), group L-V: LDS ceramic (veneering), group ZL-M: zirconia-reinforced lithium silicate ceramic (ZLS, monolithic), group ZL-V: ZLS ceramic (veneering). Crown restorations were subjected to load-to-failure test (0.5 mm/min). Failure loads of each group were statistically analyzed (two-way ANOVA, post hoc Tukey HSD, alpha = 0.05). Finite element analysis (FEA) was used to compare the stress distribution of crown restorations. Results Group L-M had the highest failure load (2891.88 +/- 410.12 N) with a significant difference from other groups (p < 0.05). Although there was a significant difference between group ZL-M (1750.28 +/- 314.96 N) and ZL-V (2202.55 +/- 503.14 N), there was no significant difference from group L-V in both groups (2077.37 +/- 356.59 N) (p > 0.05). Conclusions The veneer application had opposite effects on ceramics, increased the failure load of ZLS and reduced it for LDS without a statistically significant difference. Both materials are suitable for implant-supported crowns. Different restorative materials did not influence the stress distribution, but monolithic restorations reduced the stress concentration on the implant and bone. | |
| dc.description.sponsorship | Abant Izzet Baysal University Scientific Research Project Department in Bolu, Turkey [BAP-2016.06.06.1039] | |
| dc.description.sponsorship | The work was supported by Abant Izzet Baysal University Scientific Research Project Department in Bolu, Turkey (project no BAP-2016.06.06.1039). | |
| dc.identifier.doi | 10.1186/s40729-020-00267-x | |
| dc.identifier.issn | 2198-4034 | |
| dc.identifier.issue | 1 | |
| dc.identifier.pmid | 33106916 | |
| dc.identifier.scopusquality | Q2 | |
| dc.identifier.uri | https://doi.org/10.1186/s40729-020-00267-x | |
| dc.identifier.uri | https://hdl.handle.net/20.500.14730/8519 | |
| dc.identifier.volume | 6 | |
| dc.identifier.wos | WOS:000584385600001 | |
| dc.identifier.wosquality | Q1 | |
| dc.indekslendigikaynak | Web of Science | |
| dc.indekslendigikaynak | PubMed | |
| dc.language.iso | en | |
| dc.publisher | Springer Japan Kk | |
| dc.relation.ispartof | International Journal of Implant Dentistry | |
| dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | |
| dc.rights | info:eu-repo/semantics/openAccess | |
| dc.snmz | KA_WOS_20250302 | |
| dc.subject | CAD | |
| dc.subject | CAM | |
| dc.subject | Ceramics | |
| dc.subject | Dental implant | |
| dc.subject | Finite element analysis | |
| dc.subject | Load-to-failure test | |
| dc.title | Comparison of CAD/CAM manufactured implant-supported crowns with different analyses | |
| dc.type | Article |
Dosyalar
Orijinal paket