Comparative performance analysis of urised 3 and DIRUI FUS-200 automated urine sediment analyzers and manual microscopic method

dc.contributor.authorYalcinkaya, Emre
dc.contributor.authorErman, Hayriye
dc.contributor.authorKirac, Eray
dc.contributor.authorSerifoglu, Afife
dc.contributor.authorAksoy, Alperen
dc.contributor.authorIsman, Ferruh K.
dc.contributor.authorCekmen, Mustafa B.
dc.date.accessioned2025-05-10T15:24:34Z
dc.date.issued2019
dc.departmentİstanbul Medeniyet Üniversitesi
dc.description.abstractObjective: Microscopic examination of urine sediment is necessary for evaluation of renal and urinary tract diseases. In this study, we evaluated and compared analytic and diagnostic performances of DIRUI FUS-200 and a new image-based automated urine sediment analyzer Urised 3. Method: A total of 440 urine samples, submitted to our laboratory, were evaluated by two automated urine sediment analyzers and a standardized manual microscopy. Precision, linearity and method comparison studies were performed according to CLSI guidelines. Results: Considering the red blood cell (RBC) and white blood cell (WBC) counts, strong correlations existed between FUS-200 and manual microscopy (r=0.993 vs 0.861), Urised 3 and manual microscopy (r=0.962 vs 0.818), FUS200 and Urised 3 (r=0.961 vs 0.961). Clinical non-concordance ranged from 7% to 14.16% among all methods. Conclusions: The concordance between the analyzers and manual microscopy for WBC was better than that of RBC. The concordance between the two analyzers was better for WBC and RBC, with respect to the manual microscopy. Although the Urised 3, FUS-200 and manual microscopy counts were in agreement; confirmation of the results of automated analyzers with manual microscopy is particularly helpful, for pathological samples with near cut-off values. © Istanbul Medeniyet University.
dc.identifier.doi10.5222/MMJ.2019.23169
dc.identifier.endpage251
dc.identifier.issn2149-2042
dc.identifier.issue3
dc.identifier.scopus2-s2.0-85073761002
dc.identifier.scopusqualityQ2
dc.identifier.startpage244
dc.identifier.trdizinid375650
dc.identifier.urihttps://doi.org/10.5222/MMJ.2019.23169
dc.identifier.urihttps://search.trdizin.gov.tr/tr/yayin/detay/375650
dc.identifier.urihttps://hdl.handle.net/20.500.14730/6790
dc.identifier.volume34
dc.indekslendigikaynakScopus
dc.indekslendigikaynakTR-Dizin
dc.language.isoen
dc.publisherLogos Medical Publishing
dc.relation.ispartofMedeniyet Medical Journal
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/openAccess
dc.snmzKA_Scopus_20250302
dc.subjectAutomated urine sediment analyzer; Comparative performance analysis; DIRUI FUS-200; Manual urine sediment analysis; Urised 3
dc.titleComparative performance analysis of urised 3 and DIRUI FUS-200 automated urine sediment analyzers and manual microscopic method
dc.title.alternativeUrised 3 ve DIRUI FUS-200 otomatik idrar sediment analizörlerinin karşılaştırmalı performans analizi ve manuel mikroskopik yöntem
dc.typeArticle

Dosyalar

Orijinal paket

Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
6790.pdf
Boyut:
949.24 KB
Biçim:
Adobe Portable Document Format