dc.description.abstractWe aimed to compare the measurements of central corneal thickness (CCT) and endothelial parameters with three different non-contact specular microscopy (SM) devices. Fifteen eyes of 15 healthy individuals (6 males; 9 females) were enrolled in the study. Mean age was 37.93 +/- 15.13 years. Endothelial parameters and CCT were measured with Nidek CEM-530, Topcon SP-3000P, and Tomey EM-3000 SM devices by the same physician. Endothelial parameters included endothelial cell count (ECC), maximum, minimum, and average endothelial cell size. and hexagonality ratio. There were no statistically significant differences in ECC, CTT, and average endothelial size (AES) between the devices (p > 0.05). The measurement of maximum endothelial size (MES) was different between Nidek SM and Topcon SM devices (p = 0.001), but there was no difference in MES between Nidek SM and Tomey SM (p = 0.058), and between Topcon SM and Tomey SM (p = 0.081). There was no difference in minimum endothelial size (MinES) between Nidek SM and Topcon SM (p = 0.794); however, there was a significant difference in MinES between Tomey SM and Nidek SM (p < 0.001), and between Tomey SM and Topcon SM (p < 0.001). Comparison of hexagonality ratio showed statistically significant difference between the devices (p < 0.001). No significant differences in the measurements of ECC, CCT, and AES were detected between different SM devices, whereas a statistically significant difference in hexagonality ratio was detected between the devices. These devices should not be used alternatively in the endothelial morphology assessment in patient's follow-up.
dc.identifier.doi10.1007/s10792-016-0264-x
dc.identifier.endpage233
dc.identifier.issn0165-5701
dc.identifier.issn1573-2630
dc.identifier.issue1
dc.identifier.pmid27221264
dc.identifier.scopus2-s2.0-84970038588
dc.identifier.scopusqualityQ2
dc.identifier.startpage229
dc.identifier.urihttps://doi.org/10.1007/s10792-016-0264-x
dc.identifier.urihttps://hdl.handle.net/20.500.14730/13251
dc.identifier.volume37
dc.identifier.wosWOS:000394159800033
dc.identifier.wosqualityQ3
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.indekslendigikaynakPubMed
dc.language.isoen
dc.publisherSpringer
dc.relation.ispartofInternational Ophthalmology
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.snmzKA_WOS_20250302
dc.subjectCorneal thickness
dc.subjectEndothelial cell count
dc.subjectEndothelial cell size
dc.subjectEndothelial parameters
dc.subjectHexagonality
dc.subjectSpecular microscopy
dc.titleMeasurements of central corneal thickness and endothelial parameters with three different non-contact specular microscopy devices
dc.typeArticle

Dosyalar

Orijinal paket

Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
13251
Boyut:
326.66 KB
Biçim:
Adobe Portable Document Format