Comparing the novel microstream and the traditional mainstream method of end-tidal CO2 monitoring with respect to PaCO2 as gold standard in intubated critically ill children

dc.authorid0000-0002-3315-6664
dc.authorid0000-0001-7892-2927
dc.authorid0000-0001-9655-3483
dc.authorid0000-0002-9609-478X
dc.authorid0000-0002-1412-0068
dc.contributor.authorDuyu, Muhterem
dc.contributor.authorBektas, Anil Dogan
dc.contributor.authorKarakaya, Zeynep
dc.contributor.authorBahar, Meral
dc.contributor.authorGunalp, Aybuke
dc.contributor.authorCaglar, Yasemin Mocan
dc.contributor.authorYersel, Meryem Nihal
dc.date.accessioned2025-05-10T19:44:22Z
dc.date.issued2020
dc.departmentİstanbul Medeniyet Üniversitesi
dc.description.abstractThe objective of this study was to evaluate a novel microstream method by comparison with PaCO2 and the more standard mainstream capnometer in intubated pediatric patients. We hypothesized that the novel microstream method would superior compared to the traditional mainstream method in predicting PaCO2. This was a prospective single-center comparative study. The study was carried out on 174 subjects with a total of 1338 values for each method. Data were collected prospectively from mainstream and microstream capnometer simultaneously and compared with PaCO2 results. Although both mainstream PetCO(2) (mainPetCO(2)) and microstream PetCO(2) (microPetCO(2)) were moderately correlated (r=0.63 and r=0.68, respectively) with PaCO2 values, mainPetCO(2) was in better agreement with PaCO2 in all subjects (bias +/- precision values of 3.8 +/- 8.9 and 7.3 +/- 8.2 mmHg, respectively). In those with severe pulmonary disease, the mainPetCO(2) and microPetCO(2) methods were highly correlated with PaCO2 (r=0.80 and r=0.81, respectively); however, the biases of both methods increased (14.8 +/- 9.1 mmHg and 16.2 +/- 9.0 mmHg, respectively). In cases with increased physiologic dead space ventilation, the agreement levels of mainPetCO(2) and microPetCO(2) methods became distorted (bias +/- precision values of 20.9 +/- 11.2 and 25.0 +/- 11.8 mm Hg, respectively) even though mainPetCO(2) and microPetCO(2) were highly correlated (r=0.78 and r=0.78, respectively). It was found that the novel microstream capnometer method for PetCO(2) measurements provided no superiority to the traditional mainstream method. Both capnometer methods may be useful in predicting the trend of PaCO2 due to significant correlations with the gold standard measurement in cases with severe pulmonary disease or increased physiological dead space -despite reduced accuracy.
dc.identifier.doi10.1038/s41598-020-79054-y
dc.identifier.issn2045-2322
dc.identifier.issue1
dc.identifier.pmid33328527
dc.identifier.scopus2-s2.0-85097618582
dc.identifier.scopusqualityQ1
dc.identifier.urihttps://doi.org/10.1038/s41598-020-79054-y
dc.identifier.urihttps://hdl.handle.net/20.500.14730/10892
dc.identifier.volume10
dc.identifier.wosWOS:000603657800008
dc.identifier.wosqualityQ1
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.indekslendigikaynakPubMed
dc.language.isoen
dc.publisherNature Portfolio
dc.relation.ispartofScientific Reports
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/openAccess
dc.snmzKA_WOS_20250302
dc.subjectArterial Carbon-Dioxide
dc.subjectBirth-Weight Infants
dc.subjectPco2 Measurements
dc.subjectEndotracheal-Tube
dc.subjectCapnography
dc.subjectSidestream
dc.subjectDifference
dc.subjectCorrelate
dc.subjectTension
dc.subjectDistal
dc.titleComparing the novel microstream and the traditional mainstream method of end-tidal CO2 monitoring with respect to PaCO2 as gold standard in intubated critically ill children
dc.typeArticle

Dosyalar

Orijinal paket

Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
10892.pdf
Boyut:
1.95 MB
Biçim:
Adobe Portable Document Format